Applications
- Pharmaceutical Industry
- Microbiological Culture Research
- Cell Culture Research
- Industrial Products
DES & DTS Series - Large Capacity Cleanroom Oven
DES830 / DTS830
Laboratory Oven is essential for general heating routines from drying food samples to curing microchips. Lab Oven is one of the fundamental instruments every laboratory should have.
DES & DTS Series is a Space saving, large capacity forced convection Class 100 clean oven.
Improved visibility and operability of control panel.
Stable cleanliness through forced circulation with rear exhaust.
Displays power consumption, CO2 emissions and heater manipulated variables on the control panel.
Incorporates a maximum of 99 steps, 99 patterns program controller with repeat function.
Offers several options such as recorder, manual/auto damper, N2 gas introducer with flow meter and emergency switch.
DES830 convertible to high performance filter type maintaining Class 100 at stable and fluctuating temperature up to a maximum of 200°C.
Stable cleanliness through forced circulation with rear exhaust.
Displays power consumption, CO2 emissions and heater manipulated variables on the control panel.
Incorporates a maximum of 99 steps, 99 patterns program controller with repeat function.
Offers several options such as recorder, manual/auto damper, N2 gas introducer with flow meter and emergency switch.
DES830 convertible to high performance filter type maintaining Class 100 at stable and fluctuating temperature up to a maximum of 200°C.
Safety features:
Enhanced safety with its phase-reversal relay detecting incorrect power source at installation.
Technical specifications
DES830 | DTS830 | ||||
System | Forced air circulation | ||||
Operating temperature range | Room temperature +30 to 260°C | Room temperature +30 to 360°C | |||
Temperature control accuracy | ±0.5°C at 260°C | ±0.5°C at 360°C | |||
Temp. fluctuation | ±0.5°C at 260°C | ±0.5°C at 360°C | |||
Temp. distribution accuracy | ±2.0°C at 260°C | ±5.0°C at 360°C | |||
Temp. gradient | ±6.0°C at 260°C | ±10.0°C at 360°C | |||
Max. temp. reaching time | ~70 mins. | ~80 mins. | |||
Interior | Stainless steel | ||||
Exterior | Cold rolled steel plate with melamine resin baking finish | ||||
Cable hole | 33mm×1 (right side) | ||||
Exhaust port | 61mm | ||||
Differential pressure meter | Analog type (0 ~ 300Pa) | ||||
Heater | 6.0kW (stainless steel pipe heater) | 9.0kW (stainless steel pipe heater) | |||
Heat insulation material | Glass wool | ||||
Heater Circuit Control | Triac zero-cross control | ||||
Fan Type / Fan Motor | Scirocco fan, condenser type motor 200W x 2 | ||||
Door | Single swing (left side) | ||||
Door switch | Door open: fan motor and heater circuit off, Door close: fan motor and heater circuit on | ||||
Filter | Heat resistant HEPA filter (dust-collection efficiency >99.97% up to 0.3μm particle filtering) | ||||
Caster wheels / adjuster | Free swivel caster wheels without stopper / level adjuster (2 at front) | ||||
Sensor | K-thermocouple double sensor (for temperature control and independent overheat prevention device) | ||||
Temperature control | PID V control by microprocessor | ||||
Temperature setting | Digital setting with ▲▼ keys | ||||
Temperature display | Green 4 digit LED digital display (top screen) and orange 5 digit LED digital display (bottom screen) (1°C resolution) | ||||
Other indications | LED indicates temperature patterns for heating/stabilizing/cooling | ||||
Timer / timer resolution | 1 min. and 99 Hrs. 59 mins: duration operation, 24 hour setting: time operation | ||||
Operation function | Constant temperature operation, Programmed operation (Maximum 99 steps, up to 99 patterns, repeat operation function), Duration/time select operation function (auto start/auto stop/quick auto stop, program operation) | ||||
Additional functions | Variable fan speed, Accumulated on time, operation time function (up to 65,535 hours); calibration offset; accumulated power consumption monitoring, total CO2 emission monitoring, heater output monitoring; power recovery; setting data save and restore | ||||
Safety device | Self-diagnostic functions (Detection for Temp. Sensor Failure, Triac Short Circuit, Automatic overheating prevention, Fan motor failure detection, Heater Line Disconnect, Main Relay Contact Damage), Earth leakage breaker, Key Lock Function, Independent overheating prevention device, Phase reversal relay, Door switch | ||||
Earth leakage breaker | 30A | 40A | |||
Leak Current/Short Circuit/Over-current Protection, Rated Current Sensitivity 30mA | |||||
Internal dimensions WxDxH | 620×480×1100 mm | ||||
External dimensions WxDxH | 850×1080×1955 mm | ||||
Internal capacity | 327L | ||||
Shelf load capacity | ~30 kg/pc. | ||||
Shelf rest step number / pitch | 35 steps / 30 mm | ||||
Power supply (50/60 Hz) | 16A (no plug, round terminal) | 24A (no plug, round terminal) | |||
Weight | ~335kg | ||||
Shelfs and Brackets (stainless) | 3 pcs. / 6 pcs. |